OptiFiber Pro

Specifikace

  Multimode module Singlemode module Quad module
Wavelengths 850 nm +/- 10 nm
1300 nm +35/-15 nm
1310 nm +/- 25 nm
1550 nm +/- 30 nm
850 nm +/- 10 nm
1300 nm +35/-15 nm
1310 nm +/- 25 nm
1550 nm +/- 30 nm
Compatible fiber types 50/125 µm
62.5/125 µm
Singlemode 50/125 µm
62.5/125 µm
Singlemode
Event dead zone 1 850 nm: 0.5 m typical
1300 nm: 0.7 m typical
1310 nm: 0.6 m typical
1550 nm: 0.6 m typical
850 nm: 0.5 m typical
1300 nm: 0.7 m typical
1310 nm: 0.6 m typical
1550 nm: 0.6 m typical
Attenuation dead zone 2 850 nm: 2.2 m typical
1300 nm: 4.5 m typical
1310 nm: 3.6 m typical
1550 nm: 3.7 m typical
850 nm: 2.2 m typical
1300 nm: 4.5 m typical
1310 nm: 3.6 m typical
1550 nm: 3.7 m typical
Dynamic range 3, 5, 6 850 nm: 28 dB typical
1300 nm: 30 dB typical
1310 nm: 32 dB typical
1550 nm: 30 dB typical
850 nm: 28 dB typical
1300 nm: 30 dB typical
1310 nm: 32 dB typical
1550 nm: 30 dB typical
Max distance range setting 40 km 130 km MM: 40 km
SM: 130 km
Distance measurement range 4, 5, 7, 8, 9, 10 850 nm: 9 km
1300 nm: 35 km
1310 nm: 80 km
1550 nm: 130 km
850 nm: 9 km
1300 nm: 35 km
1310 nm: 80 km
1550 nm: 130 km
Reflectance range 4, 5 850 nm: -14 dB to -57 dB (typical)
1300 nm: -14 dB to -62 dB (typical)
1310 nm: -14 dB to -65 dB (typical)
1550 nm: -14 dB to -65 dB (typical)
850 nm: -14 dB to -57 dB (typical)
1300 nm: -14 dB to -62 dB (typical)
1310 nm: -14 dB to -65dB (typical)
1550 nm: -14 dB to -65 dB (typical)
Sample resolution 3 cm to 400 cm 3 cm to 400 cm 3 cm to 400 cm
Pulse widths (nominal) 850 nm: 3, 5, 20, 40, 200 ns
1300 nm: 3, 5, 20, 40, 200, 1000 ns
1310/1550 nm: 3, 10, 30, 100, 300, 1000, 3000, 10000, 20000 ns 850 nm: 3, 5, 20, 40, 200 ns
1300 nm: 3, 5, 20, 40, 200, 1000 ns
1310/1550 nm: 3, 10, 30, 100, 300, 1000, 3000, 10000, 20000 ns
Test time (per wavelength) Auto setting: 5 sec (typical) Auto setting: 10 sec (typical) Auto setting:
MM – 5 sec (typical)
SM – 10 sec (typical)
Quick test setting:
2 sec (typical)
Quick test setting:
5 sec (typical)
Quick test setting:
MM – 2 sec (typical)
SM – 5 sec (typical)
Best resolution setting:
2 to 180 sec
Best resolution setting:
5 to 180 sec
Best resolution setting:
MM – 2 to 180 sec M
SM – 5 to 180 sec
FaultMap setting:
2 sec (typical), 180 sec (max)
FaultMap setting:
10 sec (typical), 180 sec (max)
FaultMap setting:
MM - 2 sec (typical), MM - 180 sec (max)
SM - 10 sec (typical), SM - 180 sec (max)
DataCenter OTDR setting:
1 sec (typical at 850 nm), 7 sec (max)
DataCenter OTDR setting:
20 sec (typical), 40 sec (max)
DataCenter OTDR setting:
MM – 1 sec (typical at 850 nm) MM – 7 sec (max)
SM – 20 sec (typical) SM – 40 sec (max)
Manual setting:
3, 5, 10, 20, 40, 60, 90, 120, 180 sec
Manual setting:
3, 5, 10, 20, 40, 60, 90, 120, 180 sec
Manual setting:
-3, 5, 10, 20, 40, 60, 90, 120, 180 sec SM - 3, 5, 10, 20, 40, 60, 90, 120, 180 sec

 

  1. Measured at 1.5 dB below non-saturating reflection peak with the shortest pulse width. Reflection peak < -40 dB for multimode and < - 50 dB for singlemode.
  2. Measured at +/- 0.5 dB deviation from backscatter with the shortest pulse width. Reflection peak < -40 dB for multimode and < - 50 dB for singlemode.
  3. For typical backscatter coefficient for OM1 fiber: 850: -65 dB, 1300: -72 dB.
  4. Typical backscatter and attenuation coefficients for OM2-OM4 fiber: 850 nm: -68 dB; 2.3 dB/km: 1300 nm: -76 dB; 0.6 dB/km.
  5. Typical backscatter and attenuation coefficients for OS1-OS2 fiber: 1310nm : -79 dB; 0.32 dB/km; 1550 nm: -82 dB; 0.19 dB/km.
  6. SNR=1 method, 3 minute averaging, widest pulse width.
  7. 850 = 9 km typical to find the end or 7 km typical to find a 0.1 dB event (with a maximum of 18 dB attenuation prior to the event).
  8. 1300 = 35 km typical to find the end or 30 km typical to find a 0.1 dB event (with a maximum of 18 dB attenuation prior to the event).
  9. 1310 = 80 km typical to find the end or 60km typical to find a 0.1 dB event (with a maximum of 20 dB attenuation prior to the event).
  10. 1550 = 130 km typical to find the end or 90 km typical to find a 0.1 dB event (with a maximum of 18 dB attenuation prior to the event).
  11. Does not include index of refraction error and does not include automatic event location error.
  12. dB variation per 1 dB step.
  13. Applies along the trace backscatter within the distance range in which the OTDR can find a 0.1 dB event.

 

Additional key specifications

FiberInspector probe specification
Magnification ~ 200X with OptiFiber Pro Display
Light source Blue LED
Power source TFS mainframe
Field of View (FOV) Horizontal: 425 µm
Vertical: 320 µm
Minimum detectable particle size 0.5 µm
Dimensions Approximately 6.75 in x 1.5 in (1175 mm x 35 mm) without adapter tip
Weight 200 g
Temperature range Operating: 32°F to 122°F (0 °C to +50 °C)
Storage: -4°F to +158°F (-20°C to +70°C)
Certifications CE (when used with the mainframe)
VFL specifications
On/Off control Mechanical switch and a button on the touch screen
Output power 316 µW (-5 dBm) = peak power = 1.0 mW (0 dBm)
Operating wavelength 650 nm nominal
Spectral width (RMS) ±3 nm
Output modes Continuous wave Pulsed mode (2 Hz to 3 Hz blink frequency)
Connector adapter 2.5 mm universal
Laser safety (classification) Class II CDRH Complies to EN 60825-2
General specifications
Weight Mainframe with module and battery: 3 lbs, 5 oz (1.28 kg)
Dimensions Mainframe with module and battery: 2.625 in x 5.25 in x 11.0 in ( 6.67 cm x 13.33 cm x 27.94 cm)
Battery Lithium ion battery pack, 7.2 volts
Battery life Eight hour Auto OTDR operation, dual wavelength, no video probe connected, 150 m of fiber
Charge Time  
Tester Off Four hours to charge from 10% to 90% capacity
Tester On Six hours to charge from 10% to 90% capacity with the tester on
Environmental specifications
Operating temperature* -18ºC to 45ºC
Non-operating temperature -30ºC to 60ºC
Operating altitude 4,000 m (13123 ft)
3200 m (10,500 ft) with ac adapter
Storage altitude 12,000 m
EMC EN 61326-1

* Using battery power. With AC power: 0ºC to 45ºC. Real Time Trace function used for no more than 5 minutes in a 15-minute period. Maximum ambient temperature is 35ºC for continuous use of the Real Time Trace function.
* Do not keep battery at temperatures below -20°C (-4°F) or above 50°C (122°F) for periods longer than one week to maintain battery capacity.



« Přehled | Specifikace OptiFiber Pro | Modely, volby a doplňky »


© FLUKETESTERY.CZ Všechna práva vyhrazena.

created by hotpotato – xhtml & css valid, use Mozilla – the best browser available | Fluke Networks

Informace o cookies | Informace o ochraně osobních údajů